首页> 外文期刊>Food Chemistry >Comparison of new approach of GC-HRMS (Q-Orbitrap) to GC-MS/MS (triple-quadrupole) in analyzing the pesticide residues and contaminants in complex food matrices
【24h】

Comparison of new approach of GC-HRMS (Q-Orbitrap) to GC-MS/MS (triple-quadrupole) in analyzing the pesticide residues and contaminants in complex food matrices

机译:GC-MS / MS(三重四极)在复杂食物基质中农药残留物和污染物的新方法比较GC-MS / MS(Triple-Quadrupole)

获取原文
获取原文并翻译 | 示例
           

摘要

Performances of multiresidue analysis of one hundred of pesticides and contaminants, using GC-Q-Orbitrap method in full scan mode were compared to those obtained with GC-triple-quadrupole method in multiple reaction monitoring mode. In terms of sensitivity, 86% of molecules exhibited lower limit of detection values using GC-Q-Orbitrap than using GC-triple-quadrupole. For the GC-Q-Orbitrap method, more than 85% of the pesticides and contaminants showed good recovery [70-120%] in wheat samples, with relative standard deviation values 20%. GC-Q-Orbitrap method appeared the most sensitive for most pesticides studied in wheat with limit of quantification values ranged between 0.1 mu g/kg and 4 mu g/kg. Moreover, the matrix effect was acceptable in wheat extracts for 84 molecules but strong suppression of the chromatographic signal was observed for 16 molecules for the GC-Q-Orbitrap method. The injection of unpurified wheat extracts spiked at 10 mu g/kg proved the potential of the GC-Q-Orbitrap method for use in performing high-throughput pesticide screening.
机译:将在多扫描模式中使用GC-Q-orbitrap方法的多重杀虫剂和污染物的多人分析的性能与多重反应监测模式中的GC-Triple-Quadruprole方法中获得的那些。就敏感性而言,86%的分子使用GC-Q-orbitrap表现出低于使用GC-Triple-Quadrupole的检测值的下限。对于GC-Q-orbitrap方法,超过85%的农药和污染物在小麦样品中显示出良好的恢复[70-120%],具有相对标准偏差值& 20%。 GC-Q-Orbitrap方法对于在小麦中研究的大多数农药最敏感的敏感性,范围为0.1μg/ kg和4μg/ kg。此外,在小麦提取物中可以接受基质效应84分子,但是对于GC-Q-壁图方法的16分子观察到色谱信号的强烈抑制。注射以10μmg/ kg掺入的未纯化的小麦提取物证明了GC-Q-orbitrap方法用于进行高通量农药筛选的潜力。

著录项

  • 来源
    《Food Chemistry》 |2021年第15期|129932.1-129932.10|共10页
  • 作者单位

    Normandie Univ Lab SMS EA3233 UNIROUEN FR3038 Pl Emile Blondel F-76821 Mont St Aignan France|SGS France Lab Rouen Technopole Madrillet 65 Ave Ettore Bugatti F-76801 St Etienne Du Rouvray France;

    SGS France Lab Rouen Technopole Madrillet 65 Ave Ettore Bugatti F-76801 St Etienne Du Rouvray France;

    SGS France Lab Rouen Technopole Madrillet 65 Ave Ettore Bugatti F-76801 St Etienne Du Rouvray France;

    SGS France Lab Rouen Technopole Madrillet 65 Ave Ettore Bugatti F-76801 St Etienne Du Rouvray France;

    Normandie Univ Lab SMS EA3233 UNIROUEN FR3038 Pl Emile Blondel F-76821 Mont St Aignan France;

    Normandie Univ Lab SMS EA3233 UNIROUEN FR3038 Pl Emile Blondel F-76821 Mont St Aignan France;

  • 收录信息 美国《科学引文索引》(SCI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Multiresidue pesticide analyses; GC-triple quadruple; GC-Q-Orbitrap; Matrix effect; Food matrices;

    机译:多土杀虫剂分析;GC-Triple四极;GC-Q-orbitrap;矩阵效应;食物矩阵;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号