首页> 外文期刊>European Journal of Orthopaedic Surgery & Traumatology >Bilateral fractures of alumina ceramic heads in total hip arthroplasty and subsequent ceramic re-fracture after revisional arthroplasty in sandwich ceramic cup
【24h】

Bilateral fractures of alumina ceramic heads in total hip arthroplasty and subsequent ceramic re-fracture after revisional arthroplasty in sandwich ceramic cup

机译:夹层陶瓷杯中全髋关节置换术中氧化铝陶瓷头的双侧骨折以及翻修后的随后陶瓷再骨折

获取原文
获取原文并翻译 | 示例
           

摘要

A case of bilateral fractures of both ceramic heads in total hip arthroplasty and subsequent re-fracture of ceramic component after revisional surgery with ceramic sandwich cup exchange is reported. The heads were a skirt-type and extra-long size. The fractures occurred 8 months after the operation without any trauma. The cause of the fractures might be the impingement between the neck and the ceramic liner rim. At revision surgery, the fractured ceramic heads were replaced with sandwich-type ceramic liner and heads with ball type. After 3 years of revision surgery, ceramic head and ceramic liner in left total hip arthroplasty were fractured without any trauma. The ceramic head and liner were replaced with metal head and polyethylene liner at re-revisional operation.
机译:据报道,在全髋关节置换术中,两个陶瓷头的双侧骨折以及在进行陶瓷三明治夹层杯翻修手术后,随后的陶瓷部件再次骨折。头是裙子型和超长尺寸。骨折发生在术后8个月,没有任何外伤。断裂的原因可能是颈部和陶瓷衬里边缘之间的碰撞。在翻修手术中,将断裂的陶瓷头替换为三明治型陶瓷衬里,并将球头替换为球形。经过三年的翻修手术,左全髋关节置换术中的陶瓷头和陶瓷衬垫破裂,没有任何创伤。在重新修订操作中,陶瓷头和衬里被金属头和聚乙烯衬里所取代。