Pacemaker Exposure to High-Power Microwave Ultrawideband Radiation

机译:Pacemaker暴露于高功率微波超宽带辐射

获取原文
获取原文并翻译 | 示例

摘要

One fundamental issue of the use of high-power microwave (HPM) sources is the health impact, in particular for patients who wear active implantable medical devices (AIMDs). This paper presents the effects of HPM ultrawideband (HPM/UWB) radiation on a pacemaker exposed to them, which is one aspect of the compatibility between HPM sources and AIMDs. The pacemaker was placed inside a phantom and exposed to a pulsed electromagnetic field with a rise time of the order of 100 ps and with an amplitude ranging between 350 kV/m and 1 MV/m. These electromagnetic field amplitudes are much higher than those used in classical scenarios involving HPM/UWB sources. Different configurations and programming modes of the pacemaker were tested. On the whole, the results of the experiments show the relative hardness and reliability of the pacemaker subjected to HPM/UWB radiation. A few nonpermanent disruptions occurred when the pulse repetition frequency and the exposure duration reached 100 Hz and 10 s, respectively.
机译:使用大功率微波(HPM)光源的一个基本问题是对健康的影响,特别是对于佩戴有源植入式医疗设备(AIMD)的患者。本文介绍了HPM超宽带(HPM / UWB)辐射对暴露于其上的起搏器的影响,这是HPM源与AIMD之间兼容性的一方面。将起搏器放置在体模内,并以100 ps量级的上升时间和350 kV / m至1 MV / m的幅度暴露于脉冲电磁场​​中。这些电磁场幅度远高于涉及HPM / UWB源的经典场景中使用的幅度。测试了起搏器的不同配置和编程模式。总体而言,实验结果显示了受到HPM / UWB辐射的起搏器的相对硬度和可靠性。当脉冲重复频率和暴露持续时间分别达到100 Hz和10 s时,发生了一些非永久性破坏。