首页> 外文期刊>Cureus. >Balloon-assisted Tracking of Guide Extension Catheter: A Novel Technique to Retrieve a Carotid Embolic Protection Device
【24h】

Balloon-assisted Tracking of Guide Extension Catheter: A Novel Technique to Retrieve a Carotid Embolic Protection Device

机译:气囊辅助跟踪引导延伸导管:一种检索颈动脉栓塞保护装置的新技术

获取原文
       

摘要

Embolic protection devices are used to minimize the risk of stroke by preventing the?migration of emboli during carotid artery stenting (CAS). After the successful conclusion of the CAS?procedure, these devices are meant to be retrieved. Sometimes retrieval of the filter can be difficult. This difficulty in retrieval can be due to multiple factors such as incomplete stent expansion, stent fracture, vasospasm, and vessel tortuosity causing pseudostenoses. In this case report, an under-expanded proximal carotid stent strut contributed to the filter not being retrievable in spite of maneuvers to retrieve the filter. An innovative approach was used; a coronary guide extension catheter was used to enhance support and balloon-assisted tracking of the extension catheter then permitted advancement of the retrieval device and ultimate retrieval of the filter. This technique to retrieve a carotid filter has not been previously described in the literature.
机译:栓塞保护装置用于通过防止颈动脉支架(CAS)期间栓塞的迁移来最小化中风的风险。在CAS的成功结束后,旨在检索这些装置。有时滤波器的检索可能很困难。检索的这种困难可能是由于多种因素,例如不完全支架的膨胀,支架骨折,血管痉挛和血管曲纹导致假脑抑制。在本例报告中,延展的近端颈动脉支架支柱响应于滤波器,尽管有何种操作来检索过滤器。使用了一种创新的方法;冠状动脉引导延伸导管用于增强延伸导管的支撑和球囊辅助跟踪,然后允许改进检索装置和终极检索过滤器。该技术以检索颈动脉滤波器尚未在文献中描述。