Combined Testing for Chlamydia, Gonorrhea, and Trichomonas by Use of the BD Max CT/GC/TV Assay with Genitourinary Specimen Types

机译:BD Max CT / GC / TV检测与泌尿生殖道标本类型的结合检测衣原体,淋病和滴虫

获取原文
获取外文期刊封面目录资料

摘要

The BD Max CT/GC/TV (MAX) assay is a true multiplex assay for simultaneous detection of chlamydia (CT), gonorrhea (GC), and trichomonas (TV). We evaluated assay performance for women using endocervical and vaginal swabs as well as urine specimens. A total of 1,143 women were tested for CT, GC, and TV and, subsequently, another 847 (1,990 total women) for CT and GC only, with positivity rates for CT, GC, and TV of 7.1%, 2.3%, and 13.5%, respectively. In men, the performance for CT and GC was determined using only urine specimens. TV performance was not assessed in male urine samples. Among men, 181/830 (21.8%) and 108/840 (12.9%) chlamydia and gonorrhea infections, respectively, were identified. Comparator assays included BD ProbeTec Chlamydia trachomatis Qx (CTQ)/Neisseria gonorrhoeae Qx (GCQ), Hologic Aptima Combo 2 (AC2) and Aptima TV (ATV), trichomonas microscopy, and culture. MAX CT sensitivity was 99.3% (95% confidence interval [CI], 96.1% to 99.9%), 95.7% (90.8% to 98.0%), 91.5% (85.8% to 95.1%), and 96.1% (92.2% to 98.1%) for vaginal swabs, endocervical swabs, female urine samples, and male urine samples, respectively. MAX GC sensitivity was 95.5% (84.9% to 98.7%), 95.5% (84.9% to 98.7%), 95.7% (85.5% to 99.8%), and 99.1% (94.9% to 99.8%) in the same order. MAX TV sensitivity was 96.1% (91.7% to 98.2%) for vaginal swabs, 93.4% (88.3% to 96.4%) for endocervical swabs, and 92.9% (87.8% to 96.0%) for female urine samples. Specificity for all organisms across all sample types was ≥98.6%. Performance estimates for the MAX assays were consistent with estimates calculated for the comparator assays (all P values were >0.1). The availability of a CT/GC/TV multiplexed assay on a benchtop instrument with a broad menu has the potential to facilitate local sexually transmitted infection (STI) testing at smaller laboratories and may encourage expanded screening for these highly prevalent infections.
机译:BD Max CT / GC / TV(MAX)分析是真正的多重分析,可同时检测衣原体(CT),淋病(GC)和滴虫(TV)。我们评估了使用宫颈管和阴道拭子以及尿液标本对女性的检测性能。总共对1,143名女性进行了CT,GC和TV测试,随后又对847名女性(共1,990名女性)进行了CT和GC测试,其CT,GC和TV阳性率分别为7.1%,2.3%和13.5。 %, 分别。在男性中,仅使用尿液标本确定CT和GC的性能。未在男性尿液样本中评估电视性能。在男性中,分别鉴定出衣原体和淋病感染为181/830(21.8%)和108/840(12.9%)。比较分析包括BD ProbeTec沙眼衣原体Q x (CTQ)/淋病奈瑟氏球菌Q x (GCQ),Hologic Aptima Combo 2(AC2)和Aptima TV(ATV),滴虫显微镜和文化。 MAX CT敏感性为99.3%(95%置信区间[CI],96.1%至99.9%),95.7%(90.8%至98.0%),91.5%(85.8%至95.1%)和96.1%(92.2%至98.1) %)分别用于阴道拭子,宫颈内拭子,女性尿液样本和男性尿液样本。 MAX GC灵敏度按相同顺序分别为95.5%(84.9%至98.7%),95.5%(84.9%至98.7%),95.7%(85.5%至99.8%)和99.1%(94.9%至99.8%)。阴道拭子的MAX TV敏感性为96.1%(91.7%至98.2%),宫颈管拭子的为93.4%(88.3%至96.4%),女性尿液样本为92.9%(87.8%至96.0%)。所有样品类型对所有生物的特异性均≥98.6%。 MAX分析的性能估算值与比较分析的估算值一致(所有 P 值均> 0.1)。在具有宽泛菜单的台式仪器上进行CT / GC / TV多重检测的潜力,可能有助于在较小的实验室进行局部性传播感染(STI)检测,并可能鼓励扩大对这些高度流行的感染的筛查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号