Low frequency absolute calibration of complex sensitivity of vector receivers in free-field

机译:自由场中矢量接收机复灵敏度的低频绝对校准

获取原文
获取外文期刊封面目录资料

摘要

Three-transducers spherical wave reciprocity method in free-field is demonstrated effectively for absolute calibration of the complex sensitivity of underwater sound pressure gradient of vector receiver in the frequency range 250 Hz to 4 kHz. The regularity of underwater sound pressure gradient distribution in the spherical wave, the theory of three-transducers reciprocity calibration method and the technique of complex moving weighted average (CMWA) are studied and reviewed. The VHS90 vector receiver manufactured by Hangzhou Applied Acoustics Research Institute (HAARI) is calibrated using underwater sound pressure gradient calibration facility in a 50 m×15 m×10 m anechoic tank. To verify the results of measurements, the VHS90 vector receiver is also calibrated using low frequency vector receiver calibration facility and the underwater sound pressure calibration facility. The calibration results and the comparisons with these facilities prove the accuracy of the calibration method and facilities described in this paper. The max deviation of modulus of complex sensitivity is 0.7 dB and max deviation of phase congruency of three channels is 1.6°.
机译:有效地证明了在自由场中使用三传感器球面波互易性方法可以绝对校准矢量接收器在250 Hz至4 kHz频率范围内的水下声压梯度的复灵敏度。研究并综述了水下声压梯度在球面波中的分布规律,三换能器互易性校正方法的理论以及复数移动加权平均(CMWA)技术。杭州应用声学研究所(HAARI)生产的VHS90矢量接收机在50 m×15 m×10 m的消声箱中使用水下声压梯度校准设备进行校准。为了验证测量结果,还使用低频矢量接收器校准工具和水下声压校准工具对VHS90矢量接收器进行了校准。校准结果以及与这些设备的比较证明了本文描述的校准方法和设备的准确性。复灵敏度模量的最大偏差为0.7 dB,三个通道的相位一致性的最大偏差为1.6°。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号