首页> 外文期刊>Transportation Research Procedia >Impact of Electromagnetic Fields in Transport on Active Implantable Medical Devices
【24h】

Impact of Electromagnetic Fields in Transport on Active Implantable Medical Devices

机译:运输中的电磁场对有源植入式医疗设备的影响

获取原文
获取外文期刊封面目录资料

摘要

This paper details numerical computation of induced interference voltage within a pacemaker due to exposure to electromagnetic (EM) fields present in a train. The goal is to verify whether electromagnetic field sources pose a potential threat to passengers with an implanted pacemaker. A simplified human body model with a bipolar lead configuration pacemaker was positioned in the vicinity of an electric distribution box - the source of an 50 Hz electric field influencing the pacemaker. Two positions were investigated, each placed in two different distances from the source of the electromagnetic (EM) field. The maximum calculated interference voltage in all simulated cases is within the 0,29 V/m range. This value corresponds to the input filter attenuation band and represents a safe value for train passengers with an implanted pacemaker, subject to undamaged distribution box and is in compliance with security requirements.
机译:本文详细介绍了由于暴露于列车中的电磁场而导致的起搏器内感应干扰电压的数值计算。目的是验证电磁场源是否对植入起搏器的乘客构成潜在威胁。具有双极引线配置起搏器的简化人体模型位于配电箱附近,配电箱是影响起搏器的50 Hz电场源。研究了两个位置,每个位置都与电磁场(EM)处于两个不同的距离。在所有模拟情况下,计算出的最大干扰电压均在0.29 V / m范围内。该值对应于输入滤波器的衰减带,代表带有植入式起搏器的火车乘客的安全值,该配电器不受配电箱损坏且符合安全要求。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号