首页> 外文期刊>Reproductive Health >Inter- and intra-observer variability in Sonographic measurements of the cross-sectional diameters and area of the umbilical cord and its vessels during pregnancy
【24h】

Inter- and intra-observer variability in Sonographic measurements of the cross-sectional diameters and area of the umbilical cord and its vessels during pregnancy

机译:妊娠期间超声检查中脐带及其血管的横截面直径和面积的观察者间和观察者内差异

获取原文
       

摘要

Background The purpose of the study was to evaluate inter- and intra-observer variability in sonographic measurements of the cross-sectional area of the umbilical cord and the diameters of its vessels in low-risk pregnancies of 12 to 40 weeks of gestation. Methods A prospective cross sectional study was performed in 221 pregnant women at different gestational ages. Measurements were carried out also by a second observer to evaluate inter-observer variability and repeated once again by the first observer to assess intra-observer variability. The linear correlation between the measurements (Spearman's coefficient of correlation) and their reliability through the intraclass correlation coefficient (ICC), the Cronbach's alpha coefficient and the limits of agreement proposed by Bland and Altman were evaluated. Results The results showed that inter-observer and intra-observer variability did not show any significant difference between examiners. A good linear correlation between the measurements and reliability was obtained, with values of R, ICC and Cronbach's alpha all above the standard limits. Conclusion It is possible to conclude that inter- and intra-observer variability in the measurements of the umbilical cord and its vessels was small; their reliability and agreement were good.
机译:背景技术这项研究的目的是评估在妊娠12至40周的低危妊娠中,超声检查中脐带截面积及其血管直径的观察者间和观察者内变异性。方法对221名不同胎龄孕妇进行了前瞻性横断面研究。第二观察者也进行测量以评估观察者间的变异性,第一观察者再次进行测量以评估观察者内的变异性。通过类内相关系数(ICC),Cronbachα系数以及Bland和Altman提出的一致性极限,评估了测量值(斯皮尔曼相关系数)与其可靠性之间的线性相关性。结果结果表明,观察者之间和观察者内部的变异性并未显示检查者之间的任何显着差异。测量值与可靠性之间具有良好的线性相关性,R,ICC和Cronbach's alpha值均高于标准限值。结论可以得出结论,观察者之间和观察者内部在测量脐带及其血管中的变异性很小;他们的可靠性和协议性都很好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号