Assessment of CT numbers in limited and medium field-of-view scans taken using Accuitomo 170 and Veraviewepocs 3De cone-beam computed tomography scanners

机译:使用Accuitomo 170和Veraviewepocs 3De锥束计算机断层扫描仪进行的有限和中等视野扫描中的CT数量评估

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Purpose To assess the influence of anatomic location on the relationship between computed tomography (CT) number and X-ray attenuation in limited and medium field-of-view (FOV) scans. Materials and Methods Tubes containing solutions with different concentrations of K2HPO4 were placed in the tooth sockets of a human head phantom. Cone-beam computed tomography (CBCT) scans were acquired, and CT numbers of the K2HPO4 solutions were measured. The relationship between CT number and K2HPO4 concentration was examined by linear regression analyses. Then, the variation in CT number according to anatomic location was examined. Results The relationship between K2HPO4 concentration and CT number was strongly linear. The slopes of the linear regressions for the limited FOVs were almost 2-fold lower than those for the medium FOVs. The absolute CT number differed between imaging protocols and anatomic locations. Conclusion There is a strong linear relationship between X-ray attenuation and CT number. The specific imaging protocol and anatomic location of the object strongly influence this relationship.
机译:目的在有限和中等视野(FOV)扫描中评估解剖位置对计算机断层扫描(CT)数量与X射线衰减之间关系的影响。材料和方法将装有不同浓度K 2 HPO 4 溶液的试管放在人头幻影的牙槽中。进行锥形束计算机断层扫描(CBCT)扫描,并测量K 2 HPO 4 溶液的CT数。通过线性回归分析确定CT数与K 2 HPO 4 浓度之间的关系。然后,检查根据解剖位置的CT数的变化。结果K 2 HPO 4 浓度与CT数呈强线性关系。有限FOV的线性回归斜率几乎比中等FOV的斜率低2倍。 CT的绝对值在成像方案和解剖位置之间有所不同。结论X射线衰减与CT数之间存在很强的线性关系。特定的成像协议和对象的解剖位置会强烈影响这种关系。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号