首页> 外文期刊>Analytical Sciences >Simple and Micro Determination of Manganese in Serum by Graphite Furnace Atomic Absorption Spectrometry
【24h】

Simple and Micro Determination of Manganese in Serum by Graphite Furnace Atomic Absorption Spectrometry

机译:石墨炉原子吸收光谱法简单,微量测定血清中的锰

获取原文
获取外文期刊封面目录资料

摘要

A method for the determination of low level manganese in serum by graphite furnace atomic absorption spectrometry was established. Serum was diluted two-fold with Triton X-100 without deproteinization. The final concentration of the detergent was 0.2%. Manganese was measured twice by 20μl injection of sample solution by a standard addition method. For the serum manganese of 0.52ng/ml, the relative standard deviation was below 5%. Special precautions for sampling of serum and some sources of contamination are also described.
机译:建立了石墨炉原子吸收光谱法测定血清中低含量锰的方法。将血清用Triton X-100稀释两倍而不脱蛋白。洗涤剂的终浓度为0.2%。通过标准添加方法,通过注入20μl样品溶液两次测量锰。对于0.52ng / ml的血清锰,相对标准偏差低于5%。还介绍了采样血清和某些污染源的特殊预防措施。