Inaccurate Blood Pressure Readings in the Intensive Care Unit: An Observational Study

机译:重症监护病房的血压读数不准确:一项观察性研究

获取原文
       

摘要

Measuring and monitoring cerebral perfusion pressure (CPP) is important in the management of patients with certain neurological conditions. To accurately reflect blood pressure at the circle of Willis, the arterial line transducer should be leveled at the tragus. This study measured the relative distance of the transducer to the tragus in 100 intensive care unit (ICU) patients in the mixed ICU at our institution, of which 44 patients had a pressure-sensitive neurological diagnosis. For neurological patients, the average distance was 10.9 cm and for non-neurological patients, the average distance was 11.4 cm (p-value: 0.60). This suggests that the arterial line transducer was leveled at approximately the same level regardless of pathology, potentially leading to falsely elevated CPP readings in patients with pressure-sensitive neurological pathology.
机译:测量和监视脑灌注压力(CPP)在某些神经系统疾病患者的治疗中很重要。为了准确反映Willis圆处的血压,应将动脉换能器对准耳屏。这项研究在我们机构的混合ICU中测量了100名重症监护病房(ICU)患者中换能器到耳屏的相对距离,其中44位患者进行了压敏神经系统诊断。神经病患者的平均距离为10.9厘米,非神经病患者的平均距离为11.4厘米(p值:0.60)。这表明不管病理如何,动脉换能器的水平均保持在大致相同的水平,这可能导致压敏神经病理患者的CPP读数错误地升高。