首页> 外文期刊>Design & Test of Computers, IEEE >In Vitro Long-Term Performance Evaluation and Improvement in the Response Time of CMOS-Based Implantable Glucose Sensors
【24h】

In Vitro Long-Term Performance Evaluation and Improvement in the Response Time of CMOS-Based Implantable Glucose Sensors

机译:基于CMOS的植入式葡萄糖传感器的体外长期性能评估和响应时间的改进

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This article confirmed that the sensor retains its measurement capability for more than 150 days in a saline solution. In addition, an optimization of the sensor structure for obtaining an improved response time was carried out. A response time as short as that of the conventional self-monitoring of blood glucose (SMBG) technology was obtained with the optimized sensor structure.
机译:本文证实了传感器在盐溶液中可保持其测量能力超过150天。另外,为了获得改善的响应时间,对传感器结构进行了优化。通过优化的传感器结构,获得了与常规血糖自我监测(SMBG)技术一样短的响应时间。