首页> 外文期刊>Control Engineering Practice >Patient-specific optimal cooling power command for hypothermia induction by liquid ventilation
【24h】

Patient-specific optimal cooling power command for hypothermia induction by liquid ventilation

机译:通过液体通气诱导体温过低的患者特定最佳冷却功率命令

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

While it is known that total liquid ventilation can rapidly cool animal subjects and improve outcomes after cardiac arrest, the temperature control strategy of the liquid ventilator for clinical use remains unknown. This work proposes to control the cooling power of the liquid ventilator in two phases. The first phase consists in a null cooling power command in order to estimate cardiac output. During the second phase, the optimal cooling power command dedicated to the estimated cardiac output is sent to the cooling system. The simulated results in human adults allow predicting the cooling performances and the overall safety.
机译:众所周知,完全的液体通气可以迅速使动物对象降温并改善心脏骤停后的结局,但临床上使用的液体通气机的温度控制策略仍然未知。这项工作建议分两个阶段控制液体呼吸机的冷却功率。第一阶段包括空冷却功率命令,以便估计心输出量。在第二阶段期间,专用于估计的心输出量的最佳冷却功率命令被发送到冷却系统。在成年人中的模拟结果可以预测冷却性能和整体安全性。