首页> 外文期刊>Communications in Numerical Methods in Engineering >Flow features and device‐induced blood trauma in CF‐VADs under a pulsatile blood flow condition: A CFD comparative study
【24h】

Flow features and device‐induced blood trauma in CF‐VADs under a pulsatile blood flow condition: A CFD comparative study

机译:脉动血流条件下CF‐VAD的血流特征和设备诱发的血液创伤:CFD比较研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

In this study, the flow features and device-associated blood trauma in 4 clinical ventricular assist devices (VADs; 2 implantable axial VADs, 1 implantable centrifugal VAD, and 1 extracorporeal VAD) were computationally analyzed under clinically relevant pulsatile flow conditions. The 4 VADs were operated at fixed pump speed at a mean rate of 4.5 L/min. Mean pressure difference, wall shear stress, volume distribution of scalar shear stress (SSS), and shear-induced hemolysis index (HI) were derived from the flow field of each VAD and were compared. The computationally predicted mean pressure difference across the 3 implantable VADs was similar to 70 mmHg, and the extracorporeal VAD was similar to 345 mmHg, which matched well with their reported pressure-flow curves. The axial VADs had higher mean wall shear stress and SSS compared with the centrifugal VADs. However, the residence time of the centrifugal VADs was much longer compared with the axial VADs because of the large volume of the centrifugal VADs. The highest SSS was observed in one axial VAD, and the longest exposure time was observed in 1 centrifugal VAD. These 2 VADs generated the highest HI. The shear-induced HI varied as a function of flow rate within each cardiac cycle. At fixed pump speed, the HI was greatest at low flow rate due to longer exposure time to shear stress compared with at high flow rate. Subsequently, we hypothesize that to reduce the risk of blood trauma during VAD support, shear stress magnitude and exposure time need to be minimized.
机译:在这项研究中,在临床相关的脉动血流条件下,通过计算分析了4种临床心室辅助设备(VAD; 2个可植入轴向VAD,1个可植入离心VAD和1个体外VAD)的血流特征和与设备相关的血液创伤。 4个VAD以固定的泵速以4.5 L / min的平均速度运行。从每个VAD的流场中得出平均压差,壁面剪应力,标量剪应力(SSS)的体积分布和剪致溶血指数(HI),并进行比较。通过计算得出的3个可植入VAD的平均压力差约为70 mmHg,体外VAD约为345 mmHg,这与它们所报告的压力-流量曲线非常吻合。与离心式VAD相比,轴向VAD具有更高的平均壁切应力和SSS。但是,由于离心式VAD的体积较大,因此与轴向式VAD相比,离心式VAD的停留时间要长得多。在一个轴向VAD中观察到最高的SSS,在1个离心VAD中观察到最长的暴露时间。这两个VAD产生最高的HI。剪切诱导的HI随每个心动周期内流速的变化而变化。在固定泵速下,由于在高流速下的暴露时间较长,因此在低流速下HI最高。随后,我们假设为减少VAD支持期间发生血液创伤的风险,需要将剪切应力的大小和暴露时间最小化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号