首页> 外文期刊>Circuits and Systems II: Express Briefs, IEEE Transactions on >TSV Based Orthogonal Coils With High Misalignment Tolerance for Inductive Power Transfer in Biomedical Implants
【24h】

TSV Based Orthogonal Coils With High Misalignment Tolerance for Inductive Power Transfer in Biomedical Implants

机译:基于TSV的正交线圈,生物医学植入物中的电感电力传输高度耐受公差

获取原文
获取原文并翻译 | 示例
           

摘要

To improve the misalignment tolerance for inductive power transfer (IPT) systems in implantable medical devices, like retinal prosthesis, a 3-D through silicon via (TSV) based orthogonal receiving coil is proposed in this brief. The mutual inductance between the receiving coil and the transmitting coil is calculated, analyzed, and compared. Results obtained from the analytical model and finite element method show the proposed coil structure significantly enhances the inductive coupling between coils under misalignment. Furthermore, we present receiving circuits with current summing techniques to enhance the energy transfer between the receiving coil to the load. The prototype of the inductive power transfer system is developed and the experimental results demonstrate that the proposed coil structure can provide a good misalignment tolerance. A maximum variation of the power transfer efficiency (PTE) is only 18.0% under angular misalignment of 90 degrees and 23.4% under lateral misalignment of 37.5% of the transmitting coil's diameter, which indicates the proposed coil structure can meet the demand of IPT systems in retinal prosthesis applications.
机译:为了提高可植入医疗装置中的感应电力传输(IPT)系统的未对准公差,如视网膜假体,在该简要内提出了基于视网膜假体的3-D通过硅通孔(TSV)正交接收线圈。计算,分析和比较,计算接收线圈和发射线圈之间的互感。从分析模型和有限元方法获得的结果表明,所提出的线圈结构显着增强了在未对准下线圈之间的电感耦合。此外,我们用电流求和技术呈现接收电路,以增强接收线圈之间的能量传递到负载。开发了感应电力传输系统的原型,实验结果表明,所提出的线圈结构可以提供良好的错位公差。功率传递效率(PTE)的最大变化在角度未对准下仅为18.0%,在横向未对准下的37.5%的透射线圈直径下的23.4%,这表明所提出的线圈结构可以满足IPT系统的需求视网膜假体应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号