首页> 外文期刊>IEEE Transactions on Biomedical Engineering >A clinically oriented video-based system for quantification of eyelid movements
【24h】

A clinically oriented video-based system for quantification of eyelid movements

机译:基于临床的基于视频的系统,用于量化眼睑运动

获取原文
获取原文并翻译 | 示例
           

摘要

A fieldworthy system was developed to quantify the eyelid movements in clinical sites. The system consists of a home-use charge-coupled device video camera, a processing unit, and a personal computer. A white marker of 4-mm diameter and 30-mg weight is attached to the lower margin of the upper eyelid. The processing unit automatically detects the vertical displacement of the upper edge of the marker. One marker is attached to each eye so that the movements of the both eyelids are measured with one camera simultaneously. The measurement error of the system was evaluated in experiments on eight healthy subjects and eight patients with eyelid paralysis. The mean of the absolute error of peak amplitudes occurring in 2 min was 0.81 mm, with the worst error being +1.7 mm. The reproducibility of the mean peak amplitude measured on five consecutive days was within 1 mm. The mean peak amplitudes of both eyes were measured preoperatively and postoperatively for approximately three months for three patients who were operated on to remove vestibular schwannoma. The results demonstrated basic clinical utility of the system.
机译:开发了一种适用于现场的系统来量化临床部位的眼睑运动。该系统由家用电荷耦合设备摄像机,处理单元和个人计算机组成。直径为4毫米,重量为30毫克的白色标记物附着在上眼睑的下缘。处理单元自动检测标记的上边缘的垂直位移。每只眼睛都附有一个标记,以便用一台摄像机同时测量两个眼皮的运动。在八名健康受试者和八名眼睑麻痹患者的实验中评估了该系统的测量误差。 2分钟内出现的峰值幅度的绝对误差的平均值为0.81 mm,最差的误差为+1.7 mm。连续五天测得的平均峰幅度的重现性在1毫米内。对于三名接受前庭神经鞘瘤切除术的患者,术前和术后约三个月测量了两只眼睛的平均峰值幅度。结果证明了该系统的基本临床实用性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号