On the accuracy of localization achievable in fiducial-based stereoscopic image registration system using an electronic portal imaging device

机译:使用电子门禁成像设备的基于基准的立体图像配准系统中可实现的定位精度

获取原文
获取原文并翻译 | 示例
       

摘要

Portal imaging using electronic portal imaging device (EPID) is a well-established image-guided radiation therapy (IGRT) technique for external beam radiation therapy. The aims of this study are threefold; (i) to assess the accuracy of isocentre localization in the fiducial-based stereoscopic image registration, (ii) to investigate the impact of errors in the beam collimation device on stereoscopic registration, and (iii) to evaluate the intra- and inter-observer variability in stereoscopic registration. Portal images of a ball bearing phantom were acquired and stereoscopic image registrations were performed based on a point centred in the ball bearing as the surrogate for registration. Experiments were replicated by applying intentional offsets in the beam collimation device to simulate collimation errors. The accuracy of fiducial markers localization was performed by repeating the experiment using three spherical lead shots implanted in a pelvic phantom. Portal images of pelvis phantom were given to four expert users to assess the inter-observer variability in performing registration. The isocentre localization accuracy tested using ball bearing phantom was within 0.3 mm. Gravity-induced systematic errors of beam collimation device by 2 mm resulted in positioning offsets of the order of 2 mm opposing the simulated errors. Relatively large inter-portal pair projection errors ranges from 1.3 mm to 1.8 mm were observed with simulated errors in the beam collimation device. The intra-user and inter-user variabilities were observed to be 0.8 and 0.4 mm respectively. Fiducial-based stereoscopic image registration using EPID is robust for IGRT procedure.
机译:使用电子门禁成像设备(EPID)进行门禁成像是一种成熟的用于外部束放射治疗的图像引导放射治疗(IGRT)技术。这项研究的目的是三个方面。 (i)评估基于基准的立体图像配准中的等中心定位的准确性,(ii)研究光束准直设备中的误差对立体配准的影响,以及(iii)评估观察者内部和观察者之间立体配准的可变性。获取球体模型的门户图像,并基于以球体中心为中心的点作为配准的替代物进行立体图像配准。通过在光束准直装置中施加故意偏移来模拟准直误差,从而重复进行实验。基准标记定位的准确性是通过使用植入盆腔体模中的三个球形铅球重复实验来进行的。骨盆幻影的门户图像已提供给四个专家用户,以评估观察者在进行注册时的变异性。使用滚珠轴承体模测试的等中心点定位精度在0.3毫米以内。重力导致的光束准直装置的系统误差为2 mm,导致与模拟误差相反的位置偏移量约为2 mm。在光束准直装置中观察到相对较大的端口间对投影误差范围为1.3 mm至1.8 mm,并且存在模拟误差。用户内和用户间的差异分别为0.8毫米和0.4毫米。使用EPID的基于基准的立体图像配准对于IGRT程序是可靠的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号