首页> 外文期刊>Atomic Energy >CALIBRATION OF NEUTRON-MULTIPLICITY DETECTORS
【24h】

CALIBRATION OF NEUTRON-MULTIPLICITY DETECTORS

机译:中子多重检测器的校准

获取原文
获取原文并翻译 | 示例
       

摘要

A method of calibrating neutron-multiplicity detectors using standard samples, certified in terms of the effective mass of (240) Pu, is described. The proposed method has an advantage over calibration using (252) Cf samples certified in terms of the neutron flux - the systematic shift of the measurements that is associated with the approximate nature of the method is eliminated. The serviceability of the proposed method of calibration is confirmed experimentally. The accuracy of the measurements of the effective mass of (240) Pu by the neutron-multiplicity method is evaluated: the standard deviation is several tenths of a percent.
机译:描述了一种使用标准样品校准中子多重检测器的方法,该标准样品的有效质量为(240)Pu。与使用中子通量证明的(252)Cf样品进行校准相比,所提出的方法具有优势-消除了与该方法的近似性质相关的测量系统偏移。实验证明了所提出的校准方法的适用性。通过中子多重性方法评估了(240)Pu有效质量的测量准确性:标准偏差为百分之几的十分之一。