The ASU IMS 3f secondary ion microanalyzer has been converted to a multiple ion detection mass spectrometer with double focusing for three neighbouring masses. We present preliminary experimental tests of the modified instrument. For that purpose we chose sulphur isotope ratio measurements in three pyrite samples using a small triple Faraday cup assembly as a multicollector. We achieved high mass resolution for off-axis masses and an accuracy of S-32/S-34 isotopic ratio of similar to 0.2%. (c) 2006 Published by Elsevier B.V.
机译:ASU IMS 3f二次离子微分析仪已转换为对三个相邻质量进行双重聚焦的多离子检测质谱仪。我们目前对改良仪器进行初步的实验测试。为此,我们使用小型三重法拉第杯组件作为多收集器,在三个黄铁矿样品中选择了硫同位素比测量。我们实现了偏轴质量的高质量分离,并且S-32 / S-34同位素比率的精度接近0.2%。 (c)2006年由Elsevier B.V.发布