The 920 MHz NMR magnet which successfully generated its designed magnetic field on April 2001 was installed at Tsukuba Magnet Laboratory (TML) of the National Institute for Materials Science (NIMS). The magnet again generated 920 MHz without a quench, and has been in continuous operation for over one and a half years up to the present time. The cryostat, which houses the superconducting magnet, employs a pressurized superfluid helium cooling process. Its operating temperature is below 1.55 K. The consumption rates of liquid helium and liquid nitrogen are 0.98 L/h and 0.76 L/h, respectively. Liquid helium is manually supplied every week and liquid nitrogen is automatically supplied every week. Although the cryostat has experienced earthquakes more than 70 times and power failure 3 times, there has been no serious trouble related to the cryostat and the cooling operation is still continuing without interruption.
机译:920 MHz NMR磁体于2001年4月成功产生其设计磁场,该磁体安装在美国材料科学研究院(NIMS)的筑波磁体实验室(TML)上。磁铁再次产生920 MHz的频率而没有失超,并且一直持续运行了一年半以上的时间。装有超导磁体的低温恒温器采用加压的超流氦冷却过程。它的工作温度低于1.55K。液氦和液氮的消耗速率分别为0.98 L / h和0.76 L / h。每周手动提供液氦,每周自动提供液氮。尽管低温恒温器经历过70次以上的地震和3次停电,但是与低温恒温器相关的严重问题并没有发生,并且冷却操作仍在继续而不会中断。