首页> 外文期刊>Applied Physics Letters >X-ray diamond detectors with energy resolution
【24h】

X-ray diamond detectors with energy resolution

机译:具有能量分辨率的X射线金刚石探测器

获取原文
获取原文并翻译 | 示例
           

摘要

Polycrystalline diamond detectors with energy resolving capability of the impinging beam were realized and tested by using a miniature pyroelectric x-ray pulse generator. Microstrip structures were defined by photolithography aimed to reduce parasitic capacitances and to perform characterization measurements in a sandwich configuration. Leakage currents as low as 20 pA at 500 V were measured on a 270 μm thick device. Pulse height distributions were carried out around Ta Lα (8.14 keV) and Cu Kα (8.05 keV) characteristic lines of the source. Energy resolution at 200 V was found equal to 9% with an increase to 11% at 500 V. When the bias was increased to the maximum voltage the sample shows an Ohmic behavior.
机译:通过使用微型热释电X射线脉冲发生器,实现了对入射光束具有能量分辨能力的多晶金刚石探测器,并进行了测试。微带结构是通过光刻法定义的,旨在减少寄生电容并以夹心结构进行表征测量。在270μm厚的器件上测得在500 V时泄漏电流低至20 pA。在源的TaLα(8.14 keV)和CuKα(8.05 keV)特征线周围进行脉冲高度分布。发现200 V时的能量分辨率等于9%,而在500 V时增加至11%。当偏压增加到最大电压时,样品表现出欧姆特性。