Harmonic focal point axial modulation (h-FPAM) in nonlinear optical microscopy is introduced and used to enhance the axial resolution and the signal to background ratio by modulating the focal point of a beam of femtosecond pulses train along the axial direction and phase sensitively filtering the resulting signals using a lock-in amplifier. Axial resolution enhancement factor of 2.05 is acquired in 2f detection mode for the third harmonic generation microscopy of glass-oil interfaces. This technique also resolves the image interpretation problem of the fundamental harmonic FPAM technique. Moreover, the potential of this technique for axial sectioning is demonstrated by acquiring images from a red blood cell.
机译:引入了非线性光学显微镜中的谐波焦点轴向调制(h-FPAM),并通过沿轴向调制飞秒脉冲序列的光束的焦点并对相位进行敏感地滤波,从而提高了轴向分辨率和信噪比。使用锁定放大器产生的信号。在2f检测模式下,对于玻璃-油界面的三次谐波产生显微镜,可获得2.05的轴向分辨率增强因子。该技术还解决了基本谐波FPAM技术的图像解释问题。此外,通过从红细胞获取图像来证明该技术用于轴向切片的潜力。

著录项

  • 来源
    《Applied Physics Letters》 |2017年第7期|071106.1-071106.5|共5页
  • 作者单位

    Laser and Plasma Research Institute, Shahid Beheshti University, G.C., Tehran 1983969411, Iran;

    Laser and Plasma Research Institute, Shahid Beheshti University, G.C., Tehran 1983969411, Iran;

    Laser and Plasma Research Institute, Shahid Beheshti University, G.C., Tehran 1983969411, Iran;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-18 03:14:01

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号