The transfer of calibration models between near-infrared spectrometric instruments with the algorithm proposed by Shenk and Westerhaus as a standardization method is studied. This algorithm does not yield optimal results when standardization samples are used that are not of the same nature as the samples to be analyzed. In an attempt to improve the results obtained with this algorithm when one uses such standardization samples, five different modifications of the spectral intensity correction step of the algorithm are investigated. The use of locally weighted regression, which gives more weight to the standardization samples close to the same spectral intensity range where the samples to be predicted are located and less weight to the farthest samples, seems to be the most suitable one.
机译:以申克和韦斯特豪斯提出的算法为标准方法,研究了近红外光谱仪之间标定模型的传递。当使用与要分析的样品性质不同的标准化样品时,该算法不会产生最佳结果。为了改善当使用这样的标准化样本时使用该算法获得的结果,研究了该算法的光谱强度校正步骤的五种不同的修改。局部加权回归的使用似乎是最合适的方法,它可以使标准化样本的权重更大,接近要预测的样本所在的相同光谱强度范围,而权重最远的样本的权重较小。