A new calibration method for a six-port reflectometer which minimizes the effect of power measurement errors is presented. It is based on the power divider correlator to obtain estimates of the six-port calibration constants. The calibration procedure is to determine the constants that govern the operation of the six-port from the measurement of power or output ports tensions by placing five to eleven standard loads in succession to DUT. A load and four offset shorts are used as reflection standards. Computer simulation results are presented to show that the described procedure substantially reduces the overall error in the calibration constants.
机译:提出了一种六端口反射仪的新校准方法,该方法可将功率测量误差的影响降至最低。它基于功率分配器相关器来获取六端口校准常数的估计值。校准程序是通过在DUT上放置5至11个标准负载,从功率或输出端口张力的测量中确定控制6端口操作的常数。负载和四个偏置短路被用作反射标准。呈现的计算机仿真结果表明,所描述的过程大大降低了校准常数中的总体误差。