A new oscillometric method for pulse wave analysis: comparison with a common tonometric method

机译:一种新的脉搏波分析示波方法:与常规眼压法的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In the European Society of Cardiology–European Society of Hypertension guidelines of the year 2007, the consequences of arterial stiffness and wave reflection on cardiovascular mortality have a major role. But the investigators claimed the poor availability of devices/methods providing easy and widely suitable measuring of arterial wall stiffness or their surrogates like augmentation index (AIx) or aortic systolic blood pressure (aSBP). The aim of this study was the validation of a novel method determining AIx and aSBP based on an oscillometric method using a common cuff (ARCSolver) against a validated tonometric system (SphygmoCor). aSBP and AIx measured with the SphygmoCor and ARCSolver method were compared for 302 subjects. The mean age was 56 years with an s.d. of 20 years. At least two iterations were performed in each session. This resulted in 749 measurements. For aSBP the mean difference was −0.1 mm Hg with an s.d. of 3.1 mm Hg. The mean difference for AIx was 1.2% with an s.d. of 7.9%. There was no significant difference in reproducibility of AIx for both methods. The variation estimate of inter- and intraobserver measurements was 6.3% for ARCSolver and 7.5% for SphygmoCor. The ARCSolver method is a novel method determining AIx and aSBP based on an oscillometric system with a cuff. The results agree with common accepted tonometric measurements. Its application is easy and for widespread use.
机译:在2007年欧洲心脏病学会-欧洲高血压学会指南中,动脉僵硬和波反射对心血管死亡率的影响具有重要作用。但是研究人员声称设备/方法的可用性很差,无法轻松,广泛地测量动脉壁硬度或其替代指标,例如增强指数(AIx)或主动脉收缩压(aSBP)。这项研究的目的是验证一种新方法,该方法基于示波法,使用通用袖带(ARCSolver)相对于经过验证的眼压测量系统(SphygmoCor)来确定AIx和aSBP。比较了用SphygmoCor和ARCSolver方法测得的aSBP和AIx对302名受试者的影响。平均年龄为56岁, 20年每个会话中至少执行两次迭代。结果进行了749次测量。对于aSBP,平均差为-0.1 mm Hg,s.d。为3.1 mm Hg。 AIx的平均差为1.2%,标准差占7.9%。两种方法在AIx的重现性上均无显着差异。观察者间和观察者间测量值的差异估计为ARCSolver为6.3%,而SphygmoCor为7.5%。 ARCSolver方法是一种基于带有袖带的示波系统确定AIx和aSBP的新颖方法。结果与公认的眼压测量结果一致。它的应用很容易,并且可以广泛使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号