Characterizing the performance of two optical particle counters (Grimm OPC1.108 and OPC1.109) under urban aerosol conditions

机译:在城市气溶胶条件下表征两个光学粒子计数器(Grimm OPC1.108和OPC1.109)的性能

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The performance of Grimm optical particle counters (OPC, models 1.108 and 1.109) was characterized under urban aerosol conditions. Number concentrations were well correlated. The different lower cut-off diameters (0.25 and 0.3 μm) give an average difference of 23.5%. Both detect less than 10% of the total particle concentration (0.01–1 μm; Differential Mobility Analyzer), but in the respective size ranges, differences are <10%. OPC number size distributions were converted to mass concentrations using instrument-specific factors given by the manufacturer. Mass concentrations for OPC1.108 were 60% higher than for OPC1.109 and (in case of OPC1.109) much lower than those measured with an impactor in the relevant size range or a TSP filter. Using the C-factor correction suggested by the manufacturer, OPC1.109 underestimated mass concentrations by 21% (impactor) and by about 36% (TSP filter), which is in the range of comparability of co-located different mass concentration methods ().
机译:Grimm光学粒子计数器(OPC,型号1.108和1.109)在城市气溶胶条件下的性能得以表征。数字浓度具有很好的相关性。不同的下截止直径(0.25和0.3μm)的平均差异为23.5%。两者都检测不到总颗粒浓度的10%(0.01-1μm;差动迁移率分析仪),但在各自的尺寸范围内,差异小于10%。使用制造商提供的特定于仪器的因子,将OPC数量大小分布转换为质量浓度。 OPC1.108的质量浓度比OPC1.109的质量浓度高60%,并且(在OPC1.109的情况下)比在相关尺寸范围或TSP过滤器中用撞击器测量的质量浓度低得多。使用制造商建议的C因子校正,OPC1.109低估了21%(影响物)和36%(TSP过滤器)的质量浓度,这在同一位置使用不同质量浓度方法的可比较性() 。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号