The acoustic properties of a robust tissue-mimicking material based on konjac–carrageenan at ultrasound frequencies in the range 5–60 MHz are described. Acoustic properties were characterized using two methods: a broadband reflection substitution technique using a commercially available preclinical ultrasound scanner (Vevo 770, FUJIFILM VisualSonics, Toronto, ON, Canada), and a dedicated high-frequency ultrasound facility developed at the National Physical Laboratory (NPL, Teddington, UK), which employed a broadband through-transmission substitution technique. The mean speed of sound across the measured frequencies was found to be 1551.7 ± 12.7 and 1547.7 ± 3.3 m s−1, respectively. The attenuation exhibited a non-linear dependence on frequency, f (MHz), in the form of a polynomial function: 0.009787f2 + 0.2671f and 0.01024f2 + 0.3639f, respectively. The characterization of this tissue-mimicking material will provide reference data for designing phantoms for preclinical systems, which may, in certain applications such as flow phantoms, require a physically more robust tissue-mimicking material than is currently available.
机译:描述了基于魔芋-角叉菜胶的坚固的组织模仿材料在5-60 MHz范围内的超声频率下的声学特性。使用两种方法表征声学特性:使用市售的临床前超声扫描仪(Vevo 770,FUJIFILM VisualSonics,多伦多,加拿大,加拿大)使用宽带反射替代技术,以及在国家物理实验室(NPL)开发的专用高频超声设备(英国泰丁顿),它采用了宽带直通传输替代技术。测得的整个频率上的平均声速分别为1551.7±12.7和1547.7±3.3 ms -1 。衰减表现出对频率f(MHz)的非线性依赖性,形式为多项式函数:0.009787f 2 + 0.2671f和0.01024f 2 + 0.3639 f。这种模仿组织的材料的特性将为临床前系统设计体模提供参考数据,在某些应用中,例如流动体模,可能需要比当前可用的物理上更坚固的组织模仿材料。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号