Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans

机译:胸腔CT的扫描质量和入口皮肤剂量:铋乳盾和后向局部CT扫描的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives. To compare the effectiveness of the bismuth breast shield and partial CT scan in reducing entrance skin dose and to evaluate the effect of the breast shield on image quality (IQ). Methods. Nanodots were placed on an adult anthropomorphic phantom. Standard chest CT, CT with shield, and partial CT were performed. Nanodot readings and effective doses were recorded. 50 patients with chest CTs obtained both with and without breast shields were reviewed. IQ was evaluated by two radiologists and by measuring Hounsfield units (HUs) and standard deviation (SD) of HU in anterior subcutaneous region. Results. Breast shield and the partial CT scans reduced radiation to the anterior chest by 38% and 16%, respectively. Partial CT increased dose to the posterior chest by 37% and effective dose by 8%. Change in IQ in shield CT was observed in the anterior chest wall. Significant change in IQ was observed in 5/50 cases. The shield caused an increase of 20 HU (P = 0.021) and a 1.86 reduction in SD of HU (P = 0.027) in the anterior compared to posterior subcutaneous regions. Summary. Bismuth breast shield is more effective than the partial CT in reducing entrance skin dose while maintaining image quality.
机译:目标。为了比较铋乳罩和局部CT扫描在减少入口皮肤剂量方面的有效性,并评估乳罩对图像质量(IQ)的影响。方法。将纳米点放置在成人拟人模型上。进行标准胸部CT,带屏蔽的CT和部分CT。记录纳米点读数和有效剂量。回顾性分析了50例既有乳盾也有无乳盾的胸部CT患者。两名放射科医生通过测量皮下前区HU的Hounsfield单位(HU)和标准偏差(SD)来评估智商。结果。乳房防护罩和部分CT扫描分别减少了对前胸的辐射,分别降低了38%和16%。部分CT使后胸部的剂量增加了37%,有效剂量增加了8%。在前胸壁观察到盾构CT的智商变化。在5/50的病例中,智商有显着变化。与后皮下区域相比,前盾增加了20 HU(P = 0.021),HU的SD降低了1.86(P = 0.027)。摘要。铋乳盾比局部CT在减少入口皮肤剂量同时保持图像质量方面更有效。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号