首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Comparison of the BacT/Alert PF Pediatric FAN Blood Culture Bottle with the Standard Pediatric Blood Culture Bottle the Pedi-BacT
【2h】

Comparison of the BacT/Alert PF Pediatric FAN Blood Culture Bottle with the Standard Pediatric Blood Culture Bottle the Pedi-BacT

机译:BacT / Alert PF小儿FAN血培养瓶与标准小儿血培养瓶Pedi-BacT的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The performance of the BacT/Alert PF (Organon-Teknika Corp., Durham, N.C.), a new nonvented pediatric FAN blood culture bottle, was compared to that of the original pediatric bottle, the Pedi-BacT, with matched aerobic cultures obtained from two separate facilities. A total of 244 clinically significant isolates were recovered from 4,015 compliant pairs. Among the positive cultures, 170 (70%) isolates were detected in both the BacT/Alert PF and the Pedi-BacT bottles, while 47 (19%) isolates were recovered in the BacT/Alert PF bottle only and 27 (11%) isolates were recovered in the Pedi-BacT bottle only. Although isolation of specific microorganisms was comparable for the two bottles, the total number of organisms recovered by the BacT/Alert PF was greater than that by the Pedi-BacT (P = 0.0272). In addition, more organisms were recovered by the BacT/Alert PF bottle from the blood of patients receiving antimicrobial therapy (P = 0.0180). Overall time to detection was similar for the two bottles; however, a significantly decreased mean time to detection was recorded for yeast from the BacT/Alert PF bottle (22.9 h; P = 0.0001) and staphylococci from the Pedi-BacT bottle (22.5 h; P = 0.0056). One false-negative culture and five false-positive cultures occurred with the Pedi-BacT bottle, compared to one false-positive culture with the BacT/Alert PF bottle. The BacT/Alert PF bottle is a reliable blood culture bottle for pediatric blood culture specimens and may offer improved recovery of microbes from patients on antimicrobial therapy. The use of the nonvented bottle will both facilitate bottle processing and decrease expenditures for materials due to the elimination of the venting needles required for the original vented bottles.
机译:BacT / Alert PF(北卡罗莱纳州达勒姆市的Organon-Teknika公司)的性能,是一种新型的无风扇FAN血液培养瓶,与原始的儿科瓶Pedi-BacT的性能进行了比较,并获得了匹配的需氧培养两个独立的设施。从4,015对顺应性对中共回收了244种临床上重要的分离株。在阳性培养物中,在BacT / Alert PF和Pedi-BacT瓶中均检测到170个(70%)分离株,而仅在BacT / Alert PF瓶中回收了47个(19%)分离株,而27个(11%)分离株仅在Pedi-BacT瓶中回收。尽管在两个瓶子中分离出的特定微生物相当,但是BacT / Alert PF回收的生物总数比Pedi-BacT回收的总数大(P = 0.0272)。此外,通过BacT / Alert PF瓶从接受抗微生物治疗的患者血液中回收了更多生物(P = 0.0180)。这两个瓶子的总检测时间相似。但是,记录到的来自BacT / Alert PF瓶的酵母(22.9 h; P = 0.0001)和来自Pedi-BacT瓶的葡萄球菌(22.5 h; P = 0.0056)的平均检测时间显着减少。与使用BacT / Alert PF瓶的一种假阳性培养物相比,Pedi-BacT瓶发生了一种假阴性培养物和五种假阳性培养物。 BacT / Alert PF瓶是用于小儿血液培养标本的可靠血液培养瓶,可提高接受抗菌治疗的患者微生物的回收率。由于省去了原始排气瓶所需的排气针,因此使用非排气瓶既可以促进瓶的加工,又可以减少材料支出。

著录项