Terahertz technology has shown broad prospects for measuring corneal water content, which is an important parameter of ocular health. Based on terahertz time-domain spectroscopy, a new indicator named characteristic ratio (CR) of the sum of low (0.2-0.7 THz) and high (0.7-1.0 THz) frequency spectral intensities, for characterizing corneal hydration is introduced in this work. CR is calculated from the real-time reflection spectra after error elimination of ex vivo human corneal stroma samples which is collected during dehydration under natural conditions (temperature: 22.4 ± 0.3°C; humidity: 20.0 ± 3%). The corresponding relationships between CR and corneal water content are reported. Comparing the linear fitting results with the published similar study, the coefficients of variation of the fitting slope and intercept are 39.4% and 27.6% lower, respectively. This indicates that this approach has the potential to achieve corneal water content in-vivo detection in the future.
机译:太赫兹技术在测量角膜含水量方面具有广阔的前景,角膜含水量是眼睛健康的重要参数。在太赫兹时域光谱学的基础上,本工作引入了一种新的指标,称为低比率(0.2-0.7 THz)和高比率(0.7-1.0 THz)频谱强度之和的特征比(CR),用于表征角膜水化。 CR是根据自然状态(温度:22.4°C±0.3°C;湿度:20.0°C±3%)脱水过程中收集的离体人角膜基质样品的误差消除后的实时反射光谱计算得出的。报道了CR与角膜含水量之间的对应关系。将线性拟合结果与已发表的类似研究进行比较,拟合斜率和截距的变异系数分别降低了39.4%和27.6%。这表明该方法有可能在将来实现体内角膜含水量检测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号