CAD/CAM transtibial prosthetic sockets from central fabrication facilities: How accurate are they?

机译:来自中央制造设施的CAD / CAM胫骨假体插槽:它们的精度如何?

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This research compares transtibial prosthetic sockets made by central fabrication facilities with their corresponding American Academy of Orthotists and Prosthetists (AAOP) electronic shape files and assesses the central fabrication process. We ordered three different socket shapes from each of 10 manufacturers. Then we digitized the sockets using a very accurate custom mechanical digitizer. Results showed that quality varied considerably among the different manufacturers. Four of the companies consistently made sockets within +/−1.1% volume (approximately 1 sock ply) of the AAOP electronic shape file, while six other companies did not. Six of the companies showed consistent undersizing or oversizing in their sockets, which suggests a consistent calibration or manufacturing error. Other companies showed inconsistent sizing or shape distortion, a difficult problem that represents a most challenging limitation for central fabrication facilities.
机译:这项研究将中央制造设施制造的胫骨假体插槽与相应的美国矫形和假肢医师学会(AAOP)电子形状文件进行了比较,并评估了中央制造过程。我们从10个制造商中订购了三种不同的插座形状。然后,我们使用非常精确的定制机械数字化仪对插座进行了数字化处理。结果表明,不同制造商之间的质量差异很大。其中有四家公司始终将插座制作在AAOP电子形状文件的+/- 1.1%的体积内(大约1层袜子),而其他六家公司则没有。其中六家公司的插座尺寸始终一致或过大,这表明校准或制造错误始终存在。其他公司显示出不一致的尺寸或形状变形,这是一个困难的问题,代表着对中央制造设施的最大挑战。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号