首页> 美国卫生研究院文献>Journal of Diabetes Science and Technology >Development and Validation of a Rapid 13C6-Glucose Isotope Dilution UPLC-MRM Mass Spectrometry Method for Use in Determining System Accuracy and Performance of Blood Glucose Monitoring Devices
【2h】

Development and Validation of a Rapid 13C6-Glucose Isotope Dilution UPLC-MRM Mass Spectrometry Method for Use in Determining System Accuracy and Performance of Blood Glucose Monitoring Devices

机译:用于确定血糖监测设备的系统准确性和性能的快速13C6-葡萄糖同位素稀释UPLC-MRM质谱方法的开发和验证

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background:There is currently considerable discussion about the accuracy of blood glucose concentrations determined by personal blood glucose monitoring systems (BGMS). To date, the FDA has allowed new BGMS to demonstrate accuracy in reference to other glucose measurement systems that use the same or similar enzymatic-based methods to determine glucose concentration. These types of reference measurement procedures are only comparative in nature and are subject to the same potential sources of error in measurement and system perturbations as the device under evaluation. It would be ideal to have a completely orthogonal primary method that could serve as a true standard reference measurement procedure for establishing the accuracy of new BGMS.
机译:背景技术:目前,有关由个人血糖监测系统(BGMS)确定的血糖浓度准确性的讨论很多。迄今为止,FDA已允许新的BGMS参照使用相同或相似的基于酶的方法确定葡萄糖浓度的其他葡萄糖测量系统来证明其准确性。这些类型的参考测量程序本质上仅是比较性的,并且与被评估设备一样,在测量和系统扰动方面也可能遭受相同的潜在误差源。具有完全正交的主要方法,可以用作建立新BGMS准确性的真正标准参考测量程序,将是理想的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号