Diagnostic Accuracy and Visual Search Efficiency: Single 8 MP vs. Dual 5 MP Displays

机译:诊断准确性和视觉搜索效率:单个8 MP与两个5 MP显示屏

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This study compared a single 8 MP vs. dual 5 MP displays for diagnostic accuracy, reading time, number of times the readers zoomed/panned images, and visual search. Six radiologists viewed 60 mammographic cases, once on each display. A sub-set of 15 cases was viewed in a secondary study using eye-tracking. For viewing time, there was significant difference (F = 13.901, p = 0.0002), with 8 MP taking less time (62.04 vs. 68.99 s). There was no significant difference (F = 0.254, p = 0.6145) in zoom/pan use (1.94 vs. 1.89). Total number of fixations was significantly (F = 4.073, p = 0.0466) lower with 8 MP (134.47 vs. 154.29). Number of times readers scanned between images was significantly fewer (F = 10.305, p = 0.0018) with 8 MP (6.83 vs. 8.22). Time to first fixate lesion did not differ (F = 0.126, p = 0.7240). It did not take any longer to detect the lesion as a function of the display configuration. Total time spent on lesion did not differ (F = 0.097, p = 0.7567) (8.59 vs. 8.39). Overall, the single 8 MP display yielded the same diagnostic accuracy as the dual 5 MP displays. The lower resolution did not appear to influence the readers’ ability to detect and view the lesion details, as the eye-position study showed no differences in time to first fixate or total time on the lesions. Nor did the lower resolution result in significant differences in the amount of zooming and panning that the readers did while viewing the cases.
机译:这项研究比较了单个8 MP与两个5 MP显示器的诊断准确性,读取时间,阅读器缩放/平移图像的次数以及视觉搜索。六位放射科医生查看了60例乳腺X线摄影病例,每次显示一次。在第二项研究中使用眼动追踪观察了15例病例的子集。对于观看时间,有显着差异(F = 13.901,p = 0.0002),其中8 MP花费的时间更少(62.04 vs.68.99 s)。缩放/平移使用上没有显着差异(F = 0.254,p = 0.6145)(1.94对1.89)。固定总数显着降低(F = 4.073,p = 0.0466),MP降低8(134.47 vs.154.29)。读取器在图像之间扫描的次数显着减少(F between = 10.305,p = 0.0018),MP为8(6.83 vs. 8.22)。初次固定病变的时间没有差异(F = 0.126,p = 0.7240)。根据显示配置,不再需要检测病变。花在病变上的总时间没有差异(F = 0.097,p = 0.7567)(8.59 vs. 8.39)。总体而言,单个8 MP显示器的诊断准确性与双5 MP显示器相同。较低的分辨率似乎并未影响读者检测和查看病变细节的能力,因为眼位研究显示,初次注视时间或病变总时间没有差异。较低的分辨率也不会导致阅读者在查看案例时所做的缩放和平移量上的显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号