Optimization of pulsed DEER measurements for Gd-based labels: choice of operational frequencies pulse durations and positions and temperature

机译:优化基于Gd的标签的脉冲DEER测量:选择工作频率脉冲持续时间和位置以及温度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In this work, the experimental conditions and parameters necessary to optimize the long-distance (≥ 60 Å) Double Electron-Electron Resonance (DEER) measurements of biomacromolecules labeled with Gd(III) tags are analyzed. The specific parameters discussed are the temperature, microwave band, the separation between the pumping and observation frequencies, pulse train repetition rate, pulse durations and pulse positioning in the electron paramagnetic resonance spectrum. It was found that: (i) in optimized DEER measurements, the observation pulses have to be applied at the maximum of the EPR spectrum; (ii) the optimal temperature range for Ka-band measurements is 14–17 K, while in W-band the optimal temperatures are between 6–9 K; (iii) W-band is preferable to Ka-band for DEER measurements. Recent achievements and the conditions necessary for short-distance measurements (<15 Å) are also briefly discussed.
机译:在这项工作中,分析了优化带有Gd(III)标签的生物大分子的长距离(≥60Å)双电子-电子共振(DEER)测量所必需的实验条件和参数。讨论的具体参数是温度,微波波段,泵浦频率与观测频率之间的间隔,脉冲串重复率,脉冲持续时间和电子顺磁共振光谱中的脉冲位置。发现:(i)在优化的DEER测量中,必须在EPR频谱的最大值处施加观察脉冲; (ii)Ka波段测量的最佳温度范围为14-17 K,而W波段的最佳温度为6-9 K; (iii)对于DEER测量,W波段优于Ka波段。还简要讨论了最近的成就以及短距离测量(<15Å)所需的条件。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号