首页> 美国卫生研究院文献>other >Chemical Exchange Saturation Transfer MR Imaging of Articular Cartilage Glycosaminoglycans at 3T: Accuracy of B0 Field Inhomogeneity Corrections with Gradient Echo Method
【2h】

Chemical Exchange Saturation Transfer MR Imaging of Articular Cartilage Glycosaminoglycans at 3T: Accuracy of B0 Field Inhomogeneity Corrections with Gradient Echo Method

机译:在3T的关节软骨糖胺聚糖的化学交换饱和转移MR成像:梯度回波法B0场不均匀性校正的准确性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

Glycosaminoglycan Chemical Exchange Saturation Transfer (gagCEST) is an important molecular MRI methodology developed to assess changes in cartilage GAG concentrations. The correction for B0 field inhomogeneity is technically crucial in gagCEST imaging. This study evaluates the accuracy of the B0 estimation determined by the dual gradient echo method and the effect on gagCEST measurements. The results were compared with those from the commonly used z-spectrum method. Eleven knee patients and three healthy volunteers were scanned. Dual gradient echo B0 maps with different ΔTE values (1, 2, 4, 6, 8, and 10 ms) were acquired. The asymmetry of the magnetization transfer ratio at 1 ppm offset referred to the bulk water frequency, MTRasym(1 ppm), was used to quantify cartilage GAG levels. The B0 shifts for all knee patients using the z-spectrum and dual gradient echo methods are strongly correlated for all ΔTE values used (r = 0.997 to 0.786, corresponding to ΔTE = 10 to 1 ms). The corrected MTRasym(1 ppm) values using the z-spectrum method (1.34% ± 0.74%) highly agree only with those using the dual gradient echo methods with ΔTE = 10 ms (1.72% ± 0.80%; r = 0.924) and 8 ms (1.50% ± 0.82%; r = 0.712). The dual gradient echo method with longer ΔTE values (more than 6 ms) has an excellent correlation with the z-spectrum method for gagCEST imaging at 3T.
机译:糖胺聚糖化学交换饱和转移(gagCEST)是一种重要的分子MRI方法,旨在评估软骨GAG浓度的变化。 B0场不均匀性的校正在gagCEST成像中在技术上至关重要。这项研究评估了由双重梯度回波方法确定的B0估计的准确性以及对gagCEST测量的影响。将结果与常用的z光谱方法进行了比较。扫描了11名膝盖患者和3名健康志愿者。获取具有不同ΔTE值(1、2、4、6、8和10 ms)的双梯度回波B0图。相对于总水频率MTRasym(1 ppm),在偏移为1 ppm时磁化传递比率的不对称性用于量化软骨GAG含量。使用z频谱和双重梯度回波方法的所有膝关节患者的B0位移与所用的所有ΔTE值都具有高度相关性(r = 0.997至0.786,对应于ΔTE= 10至1 ms)。使用z光谱方法校正的MTRasym(1 ppm)值(1.34%±0.74%)仅与使用双梯度回波方法的ΔTE= 10 ms(1.72%±0.80%; r = 0.924)和8高度一致ms(1.50%±0.82%; r = 0.712)。 ΔTE值较长(大于6 ms)的双梯度回波方法与z光谱方法在3T下进行gagCEST成像具有良好的相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号