Effects of Inversion Time on Inversion Recovery Prepared Ultrashort Echo Time (IR-UTE) Imaging of Free and Bound Water in Cortical Bone

机译:反转时间对皮质骨中游离水和束缚水的反转恢复制备的超短回波时间(IR-UTE)成像的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Water is present in cortical bone in different binding states. In this study we aimed to investigate the effects of inversion time (TI) on the signal from bound and pore water in cortical bone using an adiabatic inversion recovery prepared ultrashort echo time (IR-UTE) sequence on a clinical 3T scanner. In total ten bovine midshaft samples and four human tibial midshaft samples were harvested for this study. Each cortical sample was imaged with the UTE and IR-UTE sequences with a TR of 300 ms and a series of TIs ranging from 10 to 240 ms. Five healthy volunteers were also imaged with the same sequence. Single- and bi-component models were utilized to calculate the T2* and relative fractions of short and long T2* components. Bi-component behavior of the signal from cortical bone was seen with the IR-UTE sequence except with a TI of around 80 ms where the short T2* component alone were seen and a mono-exponential decay pattern was observed. In vivo imaging with the IR-UTE sequence provided high contrast-to-noise images with direct visualization of bound water and reduced signal from long T2 muscle and fat. Our preliminary results demonstrate that selective nulling of the pore water component can be achieved with the IR-UTE sequence with an appropriate TI, allowing selective imaging of the bound water component in cortical bone in vivo using clinical MR scanners.
机译:水以不同的结合状态存在于皮质骨中。在这项研究中,我们旨在通过在临床3T扫描仪上使用绝热反转恢复制备的超短回波时间(IR-UTE)序列,研究反转时间(TI)对皮质骨中结合水和孔隙水信号的影响。总共收集了十个牛中轴样本和四个人类胫骨中轴样本用于这项研究。使用UTE和IR-UTE序列对每个皮质样品进行成像,TR值为300 ms,一系列TI的范围为10到240 ms。五名健康​​志愿者也以相同的顺序成像。利用单组分和双组分模型来计算T2 *和长,短T2 *组分的相对分数。用IR-UTE序列观察到来自皮质骨的信号的双组分行为,除了TI大约为80ms,其中仅观察到短的T2 *组分并且观察到单指数衰减模式。使用IR-UTE序列进行的体内成像可提供高对比度的噪声图像,并直接可视化结合水并减少来自长T2肌肉和脂肪的信号。我们的初步结果表明,使用具有适当TI的IR-UTE序列可以实现孔隙水成分的选择性清零,从而可以使用临床MR扫描仪在体内对皮质骨中结合的水分成分进行选择性成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号