首页> 美国卫生研究院文献>other >Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry
【2h】

Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry

机译:比较使用扫频光源光学相干断层扫描仪的新型生物测定仪和使用部分相干干涉法的传统生物测定仪

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aim of this study was to compare the axial lengths (ALs) using a new biometer with swept-source optical coherence tomography (Argos) versus ALs using a conventional biometer with partial coherence interferometry (IOL Master, version 5). The ALs in 48 eyes of 48 cataract patients were measured with Argos using refractive indexes that correspond to the particular tissue and with IOL Master using a single refractive index. The eyes were divided into three subgroups by AL length: short-AL group (n = 16), <23.27 mm; intermediate-AL group (n = 16), 23.27–24.03 mm; long-AL group (n = 16), ≥24.04 mm. The ALs (mm) measured with the Argos and IOL Master biometers, respectively, were 22.77 ± 0.43 and 22.74 ± 0.44, 23.63 ± 0.21 and 23.62 ± 0.21, and 26.00 ± 1.61 and 26.05 ± 1.64 in the short-, intermediate-, and long-AL groups, respectively. The mean ALs with the Argos biometer were longer than those with the IOL Master biometer in the short-AL group (P = 0.002) There was no significant difference in the intermediate-AL groups (P = 0.14). In contrast, the mean ALs with the Argos biometer were shorter than those with the IOL Master biometer in the long-AL group (P < 0.001). Differences between the ALs measured with the two biometers were statistically significant in short- and long-AL subgroups. However, the differences might not be clinically significant.
机译:这项研究的目的是比较使用带有扫频光学相干断层扫描(Argos)的新型生物测定仪与使用具有部分相干干涉法的传统生物测定仪(IOL Master,版本5)的AL的轴向长度(AL)。用与特定组织相对应的折射率,用Argos测量48位白内障患者的48只眼中的AL,并使用单一折射率,用IOL Master测量。根据AL长度将眼睛分为三个亚组:短AL组(n = 16),<23.27 mm;中度AL组(n = 16),23.27-24.03毫米;长AL组(n = 16),≥24.04mm。用Argos和IOL Master生物测定仪测得的ALs(mm)分别为22.77±0.43和22.74±0.44、23.63±0.21和23.62±0.21,以及26.00±1.61和26.05±1.64。长AL组。在短AL组中,使用Argos生物测定仪的平均ALs大于使用IOL Master生物测定仪的平均AL(P = 0.002),在中度AL组中,没有显着差异(P = 0.14)。相反,在长AL组中,使用Argos生物测定仪的平均AL短于使用IOL Master生物测定仪的平均AL(P <0.001)。用两种生物测定仪测量的AL之间的差异在短AL和长AL亚组中具有统计学意义。但是,这些差异在临床上可能并不重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号