A method is presented for checking the treatment time calculation for high dose rate (HDR) vaginal cylinder treatments. The method represents an independent check of the HDR planning system and can take into account nonuniform isodose line coverage around the cylinder. Only the air kerma strength of the source and information that is available from the written directive are required. The maximum discrepancy for a representative set of cylinder plans done on a Nucletron unit was 5%. A working HTML JavaScript program is included in the .PACS number(s): 87.53.–j, 87.90.+y
机译:提出了一种检查高剂量率(HDR)阴道圆筒治疗的治疗时间计算的方法。该方法代表对HDR规划系统的独立检查,并且可以考虑到圆柱体周围的等剂量线覆盖范围不均匀。仅要求源的空气比释动能强度和书面指令中可提供的信息。在Nucletron装置上完成的一组代表性气缸计划的最大差异为5%。一个有效的HTML JavaScript程序包含在.PACS号中:87.53.–j,87.90。+ y