Verification of the accuracy of 3D calculations of breast dose during tangential irradiation: measurements in a breast phantom

机译:验证切线辐照期间乳房剂量的3D计算准确性:乳房模型中的测量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This report specifically describes the use of a unique anthropomorphic breast phantom to validate the accuracy of three‐dimensional dose calculations performed by a commercial treatment‐planning system for intact‐breast tangential irradiation. The accuracy of monitor‐unit calculations has been corroborated using ionization chamber measurements made in this phantom. Measured doses have been compared to those calculated from a variety of treatment plans. The treatment plans utilized a 6‐MV x‐ray beam and incorporated a variety of field configurations and wedge combinations. Dose measurements at several clinically relevant points within the breast phantom have confirmed the accuracy of calculated doses generated from the variety of treatment plans. Overall agreement between measurements and calculations averaged 0.998 ± 0.009. These results indicate that the dose per monitor‐unit calculations performed by the treatment‐planning system can be confidently utilized in the fulfillment of clinical dose prescriptions.PACS number(s): 87.53.–j, 87.66.–a
机译:该报告专门描述了使用独特的拟人化乳房幻像来验证由商业治疗计划系统对完整乳房切线照射进​​行三维剂量计算的准确性。使用此模型中的电离室测量值,证实了监测器单元计算的准确性。已将测得的剂量与根据各种治疗计划计算出的剂量进行了比较。治疗计划利用了6 MV X射线束,并结合了多种现场配置和楔形组合。乳房幻影内几个临床相关点的剂量测量已经确认了各种治疗计划产生的计算剂量的准确性。测量和计算之间的总体一致性平均为0.998±0.009。这些结果表明,由治疗计划系统执行的每监测单位剂量计算可以放心地用于临床剂量处方。PACS数量:87.53.–j,87.66.–a

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号