High‐precision GAFCHROMIC EBT film‐based absolute clinical dosimetry using a standard flatbed scanner without the use of a scanner non‐uniformity correction

机译:使用标准平板扫描仪进行高精度GAFCHROMIC EBT胶片绝对临床剂量测定无需使用扫描仪不均匀性校正

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

To report a study of the use of GAFCHROMIC EBT radiochromic film (RCF) digitized with a commercially available flatbed document scanner for accurate and reliable all‐purpose two‐dimensional (2D) absolute dosimetry within a clinical environment. We used a simplified methodology that yields high‐precision dosimetry measurements without significant postirradiation correction. The Epson Expression 1680 Professional scanner and the Epson Expression 10000XL scanner were used to digitize the films. Both scanners were retrofitted with light‐diffusing glass to minimize the effects of Newton rings. Known doses were delivered to calibration films. Flat and wedge fields were irradiated with variable depth of solid water and 5 cm back scatter solid water. No particular scanner nonuniformity effect corrections or significant post‐scan image processing were carried out. The profiles were compared with CC04 ionization chamber profiles. The depth dose distribution was measured at a source‐to‐surface distance (SSD) of 100 cm with a field size of 10 × 10 cm2. Additionally, 22 IMRT fields were measured and evaluated using gamma index analysis. The overall accuracy of RCF with respect to CC04 was found to be 2%–4%. The overall accuracy of RCF was determined using the absolute mean of difference for all flat and wedge field profiles. For clinical IMRT fields, both scanners showed an overall gamma index passing rate greater than 90%. This work demonstrated that EBT films, in conjunction with a commercially available flatbed scanner, can be used as an accurate and precise absolute dosimeter. Both scanners showed that no significant scanner nonuniformity correction is necessary for accurate absolute dosimetry using the EBT films for field sizes smaller than or equal to 15 × 15 cm2.PACS number: 87.53.Bn
机译:报告关于使用GAFCHROMIC EBT放射变色胶片(RCF)进行数字化处理的研究,该胶片已通过商业化的平板文档扫描仪进行数字化,以在临床环境中进行准确可靠的通用二维(2D)绝对剂量测定。我们使用了一种简化的方法,可进行高精度的剂量学测量,而无需进行大量的辐照后校正。使用Epson Expression 1680 Professional扫描仪和Epson Expression 10000XL扫描仪将胶片数字化。两种扫描仪都用光散射玻璃进行了改装,以最大程度地减少牛顿环的影响。将已知剂量输送至校准膜。用可变深度的固体水和5厘米向后散射的固体水照射平场和楔形场。没有进行特定的扫描仪不均匀性影响校正或重大的扫描后图像处理。将轮廓与CC04电离室轮廓进行比较。在源到表面距离(SSD)为100 cm的情况下测量深度剂量分布,场大小为10×10 cm 2 。另外,使用伽马指数分析测量和评估了22个IMRT场。相对于CC04,RCF的整体精度为2%–4%。 RCF的整体精度是使用所有平坦和楔形场剖面的绝对差均值确定的。对于临床IMRT领域,这两种扫描仪均显示总体伽玛指数通过率大于90%。这项工作表明,EBT膜与市售的平板扫描仪一起可以用作准确和精确的绝对剂量计。两种扫描仪均表明,对于EBT胶片,对于小于或等于15×15 cm 2 的场尺寸,不需要进行任何显着的扫描仪不均匀校正即可进行准确的绝对剂量测定.PACS编号:87.53.Bn

著录项

相似文献

  • 外文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号