Pacing failure caused by automatic pacing threshold adjustment system

机译:自动起搏阈值调整系统导致的起搏失败

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient.
机译:心室捕获管理是一种自动起搏阈值调整算法,通过检测起搏刺激后诱发的反应来自动测量起搏阈值。尽管其主要目的是在维护患者安全的同时节省设备电池,但由于该算法的弱点,我们遇到了罕见的严重起搏失败案例。这种起搏失败的原因可能是心室起搏阈值的大幅度变化,这取决于房室间隔和起搏阈值的每日变化以及该患者的类固醇使用情况。