>Purpose: Megavoltage photon beams are typically used for therapy because of their skin-sparing effect. However, a focused low-energy x-ray beam would also be skin sparing, and would have a higher dose concentration at the focal spot. Such a beam can be produced with polycapillary optics. MCNP5 was used to model dose profiles for a scanned focused beam, using measured beam parameters. The potential of low energy focused x-ray beams for radiation therapy was assessed.>Methods: A polycapillary optic was used to focus the x-ray beam from a tungsten source. The optic was characterized and measurements were performed at 50 kV. PMMA blocks of varying thicknesses were placed between optic and the focal spot to observe any variation in the focusing of the beam after passing through the tissue-equivalent material. The measured energy spectrum was used to model the focused beam in MCNP5. A source card (SDEF) in MCNP5 was used to simulate the converging x-ray beam. Dose calculations were performed inside a breast tissue phantom.>Results: The measured focal spot size for the polycapillary optic was 0.2 mm with a depth of field of 5 mm. The measured focal spot remained unchanged through 40 mm of phantom thickness. The calculated depth dose curve inside the breast tissue showed a dose peak several centimeters below the skin with a sharp dose fall off around the focus. The percent dose falls below 10% within 5 mm of the focus. It was shown that rotating the optic during scanning would preserve the skin-sparing effect of the focused beam.>Conclusions: Low energy focused x-ray beams could be used to irradiate tumors inside soft tissue within 5 cm of the surface.
机译:>目的:兆伏光子束由于可以起到保护皮肤的作用,通常用于治疗。但是,聚焦的低能量X射线束也将节省皮肤,并且在焦点处具有更高的剂量浓度。这样的光束可以用多毛细管光学器件产生。 MCNP5用于使用测得的光束参数对扫描聚焦光束的剂量分布进行建模。评估了低能量聚焦X射线束在放射治疗中的潜力。>方法:使用多毛细管光学器件聚焦来自钨源的X射线束。对光学器件进行表征,并在50 kV下进行测量。将厚度不同的PMMA块放置在光学元件和焦点之间,以观察穿过组织等效材料后光束聚焦的任何变化。测得的能谱用于模拟MCNP5中的聚焦光束。 MCNP5中的源卡(SDEF)用于模拟会聚的X射线束。在乳房组织体模内进行剂量计算。>结果:多毛细血管镜的测量焦点尺寸为0.2 mm,景深为5 mm。在40毫米幻象厚度范围内,测得的焦点保持不变。计算出的乳房组织内部的深度剂量曲线显示,在皮肤以下几厘米处出现剂量峰值,并且在焦点周围出现了急剧的剂量下降。剂量百分比在焦点的5毫米内降至10%以下。结果表明,在扫描过程中旋转光学元件将保留聚焦光束的保留皮肤的作用。>结论:低能量聚焦X射线束可用于照射5 cm内的软组织内部的肿瘤。表面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号