2D measurements of cup orientation are less reliable than 3D measurements

机译:杯子方向的2D测量不如3D测量可靠

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background and purpose 2D analysis of metal-on-metal (MoM) hip arthroplasty (HA) has been conducted in several large series on conventional radiographs with the use of Ein Bild Roentegen Analyse (EBRA) software, but there have been no comparisons with 3D analysis in the literature. The main aim of this study was to quantify the agreement in measurements of cup version of large-diameter MoM hips obtained by EBRA and by 3D computed tomography (3D-CT). The secondary aim was to quantify the agreement for cup inclination. Lastly, we wanted to determine the inter- and intra-observer reliability of both methods.
机译:背景和目的已使用Ein Bild Roentegen Analyze(EBRA)软件在常规X射线照片上进行了多个金属对金属(MoM)髋关节置换术(HA)的2D分析,但没有与3D进行比较文献中的分析。这项研究的主要目的是量化通过EBRA和3D计算机断层扫描(3D-CT)获得的大直径MoM髋关节杯形测量的一致性。第二个目的是量化杯子倾斜度的一致性。最后,我们想确定两种方法在观察者之间和观察者内部的可靠性。