Reliability of transcutaneous monitoring of arterial Po2 in newborn infants

机译:经皮监测婴儿动脉Po2的可靠性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

Blood Po2 was measured transcutaneously on skin `arterialized' by means of a heated Po2 electrode. Oxygen diffuses through the relatively thin skin of newborn infants from the outer capillary layer to an electrode fixed to the skin. Arterial changes were registered with approximately a 10-second delay. 41 comparative transcutaneous and arterial Po2 measurements were performed. Transcutaneous Po2 values correlated quite well with arterial values within the whole Po2 range, the correlation being particularly satisfactory in the Po2 range below 100 mmHg. The method is suitable for long-term monitoring of newborn infants, as it is without risk.
机译:血液Po2是通过加热的Po2电极在“动脉化”的皮肤上经皮测量的。氧气从新生婴儿的相对较薄的皮肤中扩散,从外毛细血管层扩散到固定在皮肤上的电极。大约10秒钟的延迟记录了动脉变化。进行了41次比较的经皮和动脉Po2测量。在整个Po2范围内,经皮Po2值与动脉值具有很好的相关性,在100 mmHg以下的Po2范围内,这种相关性特别令人满意。该方法适用于新生儿的长期监测,因为它没有风险。