Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements

机译:用于骨矿物质测量的GE Healthcare月球奇才和iDXA双能X射线密度计的交叉校准

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21–72 years) underwent anteroposterior lumbar spine L2–L4 (n = 72), proximal femur (n = 72), and total body (n = 62) measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) measurements with these two scanners showed a high linear association at all tested sites (r = 0.962–0.995; p < 0.0001). The average iDXA BMD values were 1.5%, 0.5%, and 0.9% higher than those of Prodigy for lumbar spine (L2–L4) (p < 0.0001), femoral neck (p = 0.048), and total hip (p < 0.0001), respectively. Total body BMD values measured with the iDXA were −1.3% lower (p < 0.0001) than those measured with the Prodigy. For total body, lumbar spine, and femoral neck, the BMD differences as measured with these two devices were independent of subject height and weight. Linear correction equations were developed to ensure comparability of BMD measurements obtained with both DXA scanners. Importantly, use of equations from previous studies would have increased the discrepancy between these particular DXA scanners, especially at hip and at spine.
机译:在长期的前瞻性研究中,不可避免地需要更换双能X射线吸收仪(DXA)。评估新旧设备的测量之间是否存在系统差异至关重要。一组女性志愿者(21-72岁)接受了Prodigy和iDXA扫描仪在腰椎前路L2-L4(n = 72),股骨近端(n = 72)和全身(n = 62)的测量。相同的访问。用这两种扫描仪进行的骨矿物质密度(BMD)测量显示,在所有被测部位,线性相关性很高(r = 0.962–0.995; p <0.0001)。 iDXA的BMD平均值分别比Prodigy的腰椎(L2-L4)(p <0.0001),股骨颈(p = 0.048)和全髋关节(p <0.0001)高1.5%,0.5%和0.9%。 , 分别。用iDXA测量的人体总BMD值比用Prodigy测量的人体BMD值低-1.3%(p <0.0001)。对于全身,腰椎和股骨颈,使用这两种设备测量的BMD差异与受试者的身高和体重无关。开发了线性校正方程,以确保使用两个DXA扫描仪获得的BMD测量值具有可比性。重要的是,使用先前研究的公式会增加这些特定DXA扫描仪之间的差异,尤其是在臀部和脊柱处。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号