Purpose. To assess the consistency of anterior segment measurements obtained using a Sirius Scheimpflug/Placido photography-based topography system (CSO, Italy) and IOLMaster partial coherence interferometry (Carl Zeiss Meditec, Germany) in eyes with cataracts. Methods. A total of 90 eyes of 90 patients were included in this prospective study. The anterior chamber depth (ACD), keratometry (K), corneal astigmatism axis, and white to white (WTW) values were randomly measured three times with Sirius and IOLMaster. Concordance between them was assessed by calculating 95% limits of agreement (LoA). Results. The ACD and K taken with the Sirius were statistically significantly higher than that taken with the IOLMaster; however, the Sirius significantly underestimated the WTW values compared with the IOLMaster. Good agreement was found for Km and ACD measurements, with 95% LoA of −0.20 to 0.54 mm and −0.16 to 0.34 mm, respectively. Poor agreement was observed for astigmatism axis and WTW measurements, as the 95% LoA was −23.96 to 23.36° and −1.15 to 0.37 mm, respectively. Conclusion. With the exception of astigmatism axis and WTW, anterior segment measurements taken by Sirius and IOLMaster devices showed good agreement and may be used interchangeably in patients with cataracts.
机译:目的。为了评估使用基于Sirius Scheimpflug / Placido摄影的地形系统(意大利CSO)和IOLMaster部分相干干涉术(德国卡尔·蔡司Meditec)获得的前段测量结果在白内障眼中的一致性。方法。这项前瞻性研究包括90例患者的90眼。用Sirius和IOLMaster随机测量前房深度(ACD),角膜曲率(K),角膜散光轴和白色至白色(WTW)值。他们之间的一致性是通过计算95%的协议限制(LoA)来评估的。结果。从统计上看,天狼星拍摄的ACD和K明显高于IOLMaster拍摄的ACD和K。但是,与IOLMaster相比,Sirius大大低估了WTW值。对于Km和ACD测量,发现有很好的一致性,其95%LoA分别为-0.20至0.54mm和-0.16至0.34mm。观察到散光轴和WTW测量的一致性差,因为95%LoA分别为-23.96至23.36°和-1.15至0.37 mm。结论。除了散光轴和WTW外,Sirius和IOLMaster设备进行的前节测量显示出良好的一致性,可以在白内障患者中互换使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号