Revision of the NIST Standard for 223Ra: New Measurements and Review of 2008 Data

机译:修订223Ra的NIST标准:新的测量和对2008年数据的审查

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

After discovering a discrepancy in the transfer standard currently being disseminated by the National Institute of Standards and Technology (NIST), we have performed a new primary standardization of the alpha-emitter 223Ra using Live-timed Anticoincidence Counting (LTAC) and the Triple-to-Double Coincidence Ratio Method (TDCR). Additional confirmatory measurements were made with the CIEMAT-NIST efficiency tracing method (CNET) of liquid scintillation counting, integral γ-ray counting using a NaI(Tl) well counter, and several High Purity Germanium (HPGe) detectors in an attempt to understand the origin of the discrepancy and to provide a correction. The results indicate that a −9.5 % difference exists between activity values obtained using the former transfer standard relative to the new primary standardization. During one of the experiments, a 2 % difference in activity was observed between dilutions of the 223Ra master solution prepared using the composition used in the original standardization and those prepared using 1 mol·L−1 HCl. This effect appeared to be dependent on the number of dilutions or the total dilution factor to the master solution, but the magnitude was not reproducible. A new calibration factor (“K-value”) has been determined for the NIST Secondary Standard Ionization Chamber (IC “A”), thereby correcting the discrepancy between the primary and secondary standards.
机译:发现当前由美国国家标准技术研究院(NIST)传播的转移标准存在差异后,我们使用实时反符合计数对α-发射体 223 Ra进行了新的主要标准化(LTAC)和三重对偶符合率方法(TDCR)。使用CIEMAT-NIST效率示踪方法(CNET)进行液体闪烁计数,使用NaI(Tl)阱计数器进行的整体γ射线计数以及几个高纯度锗(HPGe)检测器进行了其他确认性测量,以试图了解差异的由来并提供更正。结果表明,相对于新的主要标准化,使用前一种转移标准获得的活性值之间存在-9.5%的差异。在其中一项实验中,观察到使用原始标准化组合物制备的 223 Ra主溶液稀释液与使用1mol·L-制备的稀释液之间的活性差异为2% 1 HCl。该效果似乎取决于稀释液的数量或对母液的总稀释倍数,但幅度不可重现。已为NIST二级标准电离室(IC“ A”)确定了新的校准因子(“ K值”),从而纠正了一级标准和二级标准之间的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号