The clinical evaluation of the Respi-check mask*: a new oxygen mask incorporating a breathing indicator

机译:Respi-check口罩*的临床评估:带有呼吸指示器的新型氧气面罩

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Study objective—To investigate the correlation between the Respi-check sensor and simultaneous chest auscultation in determining the respiratory rates in adults. Methods—Random visits to a local accident and emergency (A&E) department were made and all patients wearing oxygen masks were recruited into the study. The new sensor was attached to the outside of the mask. One researcher auscultated the chest to count breaths, the other counted the sensor activity. Each was blinded to the activities of the other. Breaths were counted by each researcher simultaneously and independently over one minute. A total of 40 patients were recruited into the study. A difference of more than two breaths/min compared with chest auscultation was deemed as a sensor failure. Results—The respiratory rates of 40 patients were measured. There were 28 men, 12 women. Twenty six patients were wearing an Intersurgical high concentration (flow 12l/min) mask, 14 were wearing an aerosol mask with variable venturi (flow 3–12l/min) by Medicaid. Over one minute rates determined by the two methods were the same in 28 cases (70%). It was accurate to within one breath in 37 cases (93%) and to within two breaths in 39 (98%) cases and in one case (2.5%) the sensor failed. The mean difference (mean of the differences between rates obtained from auscultation and the new sensor) was -0.1282 breaths/min, with limits of agreement (d (2SD) between -1.414 to 1.157 breaths/min. Conclusion—The Respi-check sensor provides an accurate method of estimating the respiratory rate in adult patients attending the A&E department.
机译:研究目的-研究Respi-check传感器与同时进行胸部听诊以确定成年人的呼吸频率之间的相关性。方法-随机访问当地的急症室(A&E),并招募了所有戴着氧气面罩的患者。新传感器安装在面罩的外部。一位研究员听诊胸部进行呼吸计数,另一位研究人员对传感器的活动进行计数。每个人都不知道对方的活动。每位研究人员在一分钟内同时独立地对呼吸进行计数。总共招募了40名患者进入研究。与胸部听诊相比,每分钟超过两次呼吸的差异被认为是传感器故障。结果:测量了40例患者的呼吸频率。男28,女12。 26名患者戴着Intersurgical高浓度(流量为12​​l / min)的口罩,其中14名戴着Medicaid的可变文丘里气雾面罩(流量为3–12l / min)。两种方法确定的超过一分钟的比率在28例中相同(70%)。在37例(93%)的一次呼吸内准确无误,在39例(98%)的情况下两次呼吸内准确,在1例(2.5%)内,传感器故障。平均差异(听诊与新传感器获得的速率之间的差异的平均值)为-0.1282呼吸/分钟,一致极限(d(2SD)在-1.414至1.157呼吸/分钟之间)结论— Respi-check传感器提供了一种准确的方法来评估参加急症室的成年患者的呼吸频率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号