首页> 中文期刊> 《中国介入影像与治疗学》 >血管位置及血管与声轴角度对HIFU表面消融效果影响的体模实验

血管位置及血管与声轴角度对HIFU表面消融效果影响的体模实验

         

摘要

目的 探讨靶区内血管位置及血管与声轴角度对HIFU表面消融效果的影响.方法 将56块含血管的仿组织体模分为A组(血管位置)和B组(血管角度),A组按血管中心与HIFU表面消融最深层面(顶面)距离d分为A1(d=0)、A2(d=10 mm)、A3(d=20 mm)亚组,B组按血管长轴与声轴之间夹角角度α分为B1(0~5°)、Bz(60°~65°)、B3(90°~95°)、B4(115°~120°)亚组.另设A4、B5亚组为空白对照组.测量各组血管内径、管壁厚度,观察辐照后损伤形态,计算能效因子(EEF).结果 A、B各亚组间血管内径、管壁厚度差异均无统计学意义(P均>0.05).损伤形态:A1、B1亚组靶区完全消融,无残留;余各亚组血管深侧靶区有残留.A组中A4亚组、A1亚组、A2亚组(由小到大排列)EEF两两比较差异均有统计学意义(P均<0.05),A1与A3亚组差异无统计学意义(P>0.05);B组中Bs亚组、Bz亚组、B3亚组(由小到大排列)EEF两两比较差异均有统计学意义(P均<0.05),B2亚组与B亚组、B5亚组与B1亚组的EEF差异无统计学意义(P均>0.05).结论 血管中心距HIFU表面消融最深层面10 mm、靶区内血管与声轴角度90°~95°对靶区消融影响最大,效率最低,消融时应注意调整以提高消融效率.%Objective To investigate the effect of blood vessels at different locations and angles between blood vessel and sonic axis within the target area on HIFU peripheral ablation.Methods Totally 56 phantoms containing rabbit thoracic a orta were divided into group A (different vessel locations) and B (different vessel angles).According to the distance (d)between the blood vessel center and the deepest (top) layer of the peripheral scanning,group A was divided into A1 (d=0),A2 (d=10 mm),A3 (d=20 mm) subgroups.According to the angle (α) between the vessel long axis and the acoustic axis,group B was divided into B1 (α=0-5°),B2(α=60° 65°),B3 (α=90°-95°),B4 (α=115°-120°) subgroups.A4,B5 subgroups were set as the control groups.The vessel diameter and vessel wall thickness were measured.The lesion morphology was observed and the energy efficiency factor (EEF) was calculated.Results There was no significant difference in vessel diameter and vessel wall thickness in both group A and B (all P>0.05).There was no residual area in the target area of A1,B1 subgroup,while there was residual area in the target area of the other subgroups.EEF of group A in ascending order were A4,A1,A2 subgroup,and statistically significant differences in pairwise comparisons (all P<0.05) except for A1 and A3 subgroup (P>0.05).EEF of group B in ascending order were B5,B2,B3 subgroup,respectively,and statistically significant differences in pairwise comparisons (all P<0.05) except for B2 and B4,B5 and B1 subgroup (both P>0.05).Conclusion It is 10 mm that the distance between the blood vessel center and the deepest (top) layer of the pe ripheral scanning,and 90°-95° that the angle between the vessel long axis and the acoustic axis has the greatest influence on the ablation of the target area.

著录项

  • 来源
    《中国介入影像与治疗学》 |2017年第7期|435-438|共4页
  • 作者单位

    重庆医科大学生物医学工程学院省部共建国家重点实验室培育基地—重庆市超声医学工程重点实验室重庆市生物医学工程学重点实验室重庆市微无创医学协同创新中心,重庆400016;

    重庆医科大学生物医学工程学院省部共建国家重点实验室培育基地—重庆市超声医学工程重点实验室重庆市生物医学工程学重点实验室重庆市微无创医学协同创新中心,重庆400016;

    重庆医科大学生物医学工程学院省部共建国家重点实验室培育基地—重庆市超声医学工程重点实验室重庆市生物医学工程学重点实验室重庆市微无创医学协同创新中心,重庆400016;

    重庆医科大学生物医学工程学院省部共建国家重点实验室培育基地—重庆市超声医学工程重点实验室重庆市生物医学工程学重点实验室重庆市微无创医学协同创新中心,重庆400016;

    重庆医科大学生物医学工程学院省部共建国家重点实验室培育基地—重庆市超声医学工程重点实验室重庆市生物医学工程学重点实验室重庆市微无创医学协同创新中心,重庆400016;

    重庆医科大学生物医学工程学院省部共建国家重点实验室培育基地—重庆市超声医学工程重点实验室重庆市生物医学工程学重点实验室重庆市微无创医学协同创新中心,重庆400016;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类 声疗法;肿瘤治疗学;
  • 关键词

    高强度聚焦超声消融; 血管; 体模;