摘要

This work investigates the design, system integration, optimization, and evaluation of amorphous silicon (a-Si:H) active matrix flat panel imagers (AMFPI) for bio-medical applications. Here, two hybrid active pixel sensor (H-APS) designs are introduced that improve the dynamic range while maintaining the desirable attributes of high speed and low noise readout. Also presented is a systematic approach for noise analysis of thin film transistors (TFT) and pixel circuits in which circuit analysis techniques and TFT noise models are combined to evaluate circuit noise performance. We also explore different options of system integration and present measurement results of a high fill-factor (HFF) array with segmented photodiode.
机译:这项工作研究了用于生物医学应用的非晶硅(a-Si:H)有源矩阵平板成像仪(AMFPI)的设计,系统集成,优化和评估。在此,介绍了两种混合有源像素传感器(H-APS)设计,这些设计在保持高速和低噪声读出的理想属性的同时,改善了动态范围。还介绍了一种用于薄膜晶体管(TFT)和像素电路的噪声分析的系统方法,该方法将电路分析技术和TFT噪声模型相结合来评估电路噪声性能。我们还探索了系统集成的不同选项,并给出了具有分段光电二极管的高填充因子(HFF)阵列的测量结果。